Evolving Horizons in Radiotherapy Auto-Contouring: Distilling Insights, Embracing Data-Centric Frameworks, and Moving Beyond Geometric Quantification
Authors:
Kareem A. Wahid,
Carlos E. Cardenas,
Barbara Marquez,
Tucker J. Netherton,
Benjamin H. Kann,
Laurence E. Court,
Renjie He,
Mohamed A. Naser,
Amy C. Moreno,
Clifton D. Fuller,
David Fuentes
Abstract:
Deep learning has significantly advanced the potential for automated contouring in radiotherapy planning. In this manuscript, guided by contemporary literature, we underscore three key insights: (1) High-quality training data is essential for auto-contouring algorithms; (2) Auto-contouring models demonstrate commendable performance even with limited medical image data; (3) The quantitative perform…
▽ More
Deep learning has significantly advanced the potential for automated contouring in radiotherapy planning. In this manuscript, guided by contemporary literature, we underscore three key insights: (1) High-quality training data is essential for auto-contouring algorithms; (2) Auto-contouring models demonstrate commendable performance even with limited medical image data; (3) The quantitative performance of auto-contouring is reaching a plateau. Given these insights, we emphasize the need for the radiotherapy research community to embrace data-centric approaches to further foster clinical adoption of auto-contouring technologies.
△ Less
Submitted 16 October, 2023;
originally announced October 2023.
-
Deep Learning-Based Dose Prediction for Automated, Individualized Quality Assurance of Head and Neck Radiation Therapy Plans
Authors:
Mary P. Gronberg,
Beth M. Beadle,
Adam S. Garden,
Heath Skinner,
Skylar Gay,
Tucker Netherton,
Wenhua Cao,
Carlos E. Cardenas,
Christine Chung,
David Fuentes,
Clifton D. Fuller,
Rebecca M. Howell,
Anuja Jhingran,
Tze Yee Lim,
Barbara Marquez,
Raymond Mumme,
Adenike M. Olanrewaju,
Christine B. Peterson,
Ivan Vazquez,
Thomas J. Whitaker,
Zachary Wooten,
Ming Yang,
Laurence E. Court
Abstract:
Purpose: This study aimed to use deep learning-based dose prediction to assess head and neck (HN) plan quality and identify suboptimal plans.
Methods: A total of 245 VMAT HN plans were created using RapidPlan knowledge-based planning (KBP). A subset of 112 high-quality plans was selected under the supervision of an HN radiation oncologist. We trained a 3D Dense Dilated U-Net architecture to pred…
▽ More
Purpose: This study aimed to use deep learning-based dose prediction to assess head and neck (HN) plan quality and identify suboptimal plans.
Methods: A total of 245 VMAT HN plans were created using RapidPlan knowledge-based planning (KBP). A subset of 112 high-quality plans was selected under the supervision of an HN radiation oncologist. We trained a 3D Dense Dilated U-Net architecture to predict 3-dimensional dose distributions using 3-fold cross-validation on 90 plans. Model inputs included CT images, target prescriptions, and contours for targets and organs at risk (OARs). The model's performance was assessed on the remaining 22 test plans. We then tested the application of the dose prediction model for automated review of plan quality. Dose distributions were predicted on 14 clinical plans. The predicted versus clinical OAR dose metrics were compared to flag OARs with suboptimal normal tissue sparing using a 2 Gy dose difference or 3% dose-volume threshold. OAR flags were compared to manual flags by 3 HN radiation oncologists.
Results: The predicted dose distributions were of comparable quality to the KBP plans. The differences between the predicted and KBP-planned D1%, D95%, and D99% across the targets were within -2.53%(SD=1.34%), -0.42%(SD=1.27%), and -0.12%(SD=1.97%), respectively, and the OAR mean and maximum doses were within -0.33Gy(SD=1.40Gy) and -0.96Gy(SD=2.08Gy). For the plan quality assessment study, radiation oncologists flagged 47 OARs for possible plan improvement. There was high interphysician variability; 83% of physician-flagged OARs were flagged by only one of 3 physicians. The comparative dose prediction model flagged 63 OARs, including 30 of 47 physician-flagged OARs.
Conclusion: Deep learning can predict high-quality dose distributions, which can be used as comparative dose distributions for automated, individualized assessment of HN plan quality.
△ Less
Submitted 25 April, 2023; v1 submitted 28 September, 2022;
originally announced September 2022.
-
A modular phantom and software to characterize 3D geometric distortion in MRI
Authors:
Jordan M. Slagowski,
Yao Ding,
Manik Aima,
Zhifei Wen,
Clifton D. Fuller,
Caroline Chung,
J. Matthew Debnam,
Ken-Pin Hwang,
Mo Kadbi,
Janio Szklaruk,
Jihong Wang
Abstract:
MRI offers outstanding soft tissue contrast that may reduce uncertainties in target and organ-at-risk delineation and enable online adaptive image-guided treatment. Spatial distortions resulting from non-linearities in the gradient fields and non-uniformity in the main magnetic field must be accounted for across the imaging field-of-view to prevent systematic errors during treatment delivery. This…
▽ More
MRI offers outstanding soft tissue contrast that may reduce uncertainties in target and organ-at-risk delineation and enable online adaptive image-guided treatment. Spatial distortions resulting from non-linearities in the gradient fields and non-uniformity in the main magnetic field must be accounted for across the imaging field-of-view to prevent systematic errors during treatment delivery. This work presents a modular phantom and software application to characterize geometric distortion (GD) within the large field-of-view MRI images required for radiation therapy simulation. The modular phantom is assembled from a series of rectangular foam blocks containing high-contrast fiducial markers in a known configuration. The modular phantom design facilitates transportation of the phantom between different MR scanners and MR-guided linear accelerators and allows the phantom to be adapted to fit different sized bores or coils. The phantom was evaluated using a 1.5T MR-guided linear accelerator (MR-Linac) and 1.5T and 3.0T diagnostic scanners. Performance was assessed by varying acquisition parameters to induce image distortions in a known manner. Imaging was performed using T1 and T2 weighted pulse sequences with 2D and 3D distortion correction algorithms and the receiver bandwidth (BW) varied as 250-815 Hz/pixel. Phantom set-up reproducibility was evaluated across independent set-ups. The software was validated by comparison with a non-modular phantom. Average geometric distortion was 0.94+/-0.58 mm for the MR-Linac, 0.90+/-0.53 mm for the 1.5 T scanner, and 1.15+/-0.62 mm for the 3.0T scanner, for a 400 mm diameter volume-of-interest. GD increased, as expected, with decreasing BW, and with the 2D versus 3D correction algorithm. Differences in GD attributed to phantom set-up were 0.13 mm or less. Differences in GD for the two software applications were less than 0.07 mm.
△ Less
Submitted 6 April, 2020; v1 submitted 23 March, 2020;
originally announced March 2020.
-
Cohort-based T-SSIM Visual Computing for Radiation Therapy Prediction and Exploration
Authors:
A. Wentzel,
P. Hanula,
T. Luciani,
B. Elgohari,
H. Elhalawani,
G. Canahuate,
D. Vock,
C. D. Fuller,
G. E. Marai
Abstract:
We describe a visual computing approach to radiation therapy (RT) planning, based on spatial similarity within a patient cohort. In radiotherapy for head and neck cancer treatment, dosage to organs at risk surrounding a tumor is a large cause of treatment toxicity. Along with the availability of patient repositories, this situation has lead to clinician interest in understanding and predicting RT…
▽ More
We describe a visual computing approach to radiation therapy (RT) planning, based on spatial similarity within a patient cohort. In radiotherapy for head and neck cancer treatment, dosage to organs at risk surrounding a tumor is a large cause of treatment toxicity. Along with the availability of patient repositories, this situation has lead to clinician interest in understanding and predicting RT outcomes based on previously treated similar patients. To enable this type of analysis, we introduce a novel topology-based spatial similarity measure, T-SSIM, and a predictive algorithm based on this similarity measure. We couple the algorithm with a visual steering interface that intertwines visual encodings for the spatial data and statistical results, including a novel parallel-marker encoding that is spatially aware. We report quantitative results on a cohort of 165 patients, as well as a qualitative evaluation with domain experts in radiation oncology, data management, biostatistics, and medical imaging, who are collaborating remotely.
△ Less
Submitted 10 October, 2019; v1 submitted 12 July, 2019;
originally announced July 2019.